TY - JOUR
T1 - Vascular complication can be minimized with a balloon-expandable, re-collapsible sheath in TAVR with a self-expanding bioprosthesis
AU - Abu Saleh, Walid K.
AU - Tang, Gilbert H.L.
AU - Ahmad, Hasan
AU - Cohen, Martin
AU - Undemir, Cenap
AU - Lansman, Steven L.
AU - Reyes, Manuel
AU - Barker, Colin M.
AU - Kleiman, Neal S.
AU - Reardon, Michael J.
AU - Ramlawi, Basel
PY - 2016/7/1
Y1 - 2016/7/1
N2 - Objective: This study evaluates the feasibility and safety of a balloon-expandable, re-collapsible sheath for TAVR patients, including those with small iliofemoral access (≤5.0 mm). Background: The recommended iliofemoral diameter for the CoreValve TAVR system is ≥6.0 mm, but the lowest limit has not been determined. Methods: Of 322 consecutive patients who underwent TAVR from 1/2014 to 8/2015 at two institutions, 64 underwent transfemoral CoreValve implantation, using an 11/19-French balloon-expandable, re-collapsible sheath, which has a 4.45 mm outer diameter (OD) on arterial entry, expands to 7.67 mm, then re-collapses upon removal. Valve sizing and vascular access were determined by computed tomography, and outcomes were assessed using the Valve Academic Research Consortium 2 (VARC-2) definitions. Results: Thirteen of 64 patients had a minimal iliofemoral artery luminal diameter (MLD) of ≤5.0 mm (mean 4.38+/−0.59 mm, range 3.1–5.0 mm), with vessel calcification ≤90° to 360° and tortuosity <45° to >90°. At the MLD point, the sheath-to-artery ratios, based on the fully expanded 7.67 mm OD, ranged 1.53–2.47, higher than previously reported ratios that risk vascular complications. Major comorbidities included chronic renal failure, severe chronic obstructive pulmonary disease, extreme thrombocytopenia, cirrhosis, prior cardiac surgery, poor ventricular function, and frailty. All 64 patients had TAVR with IV sedation and local anesthesia, with 0% sheath malfunction, 0% vascular complications, and 0% bleeding in-hospital and at 30 days per VARC-2 definitions. Conclusions: TAVR using a balloon-expandable, re-collapsible sheath is safe, including in small iliofemoral access ≤5.0 mm, thus considerably expanding the population suitable for transfemoral approach.
AB - Objective: This study evaluates the feasibility and safety of a balloon-expandable, re-collapsible sheath for TAVR patients, including those with small iliofemoral access (≤5.0 mm). Background: The recommended iliofemoral diameter for the CoreValve TAVR system is ≥6.0 mm, but the lowest limit has not been determined. Methods: Of 322 consecutive patients who underwent TAVR from 1/2014 to 8/2015 at two institutions, 64 underwent transfemoral CoreValve implantation, using an 11/19-French balloon-expandable, re-collapsible sheath, which has a 4.45 mm outer diameter (OD) on arterial entry, expands to 7.67 mm, then re-collapses upon removal. Valve sizing and vascular access were determined by computed tomography, and outcomes were assessed using the Valve Academic Research Consortium 2 (VARC-2) definitions. Results: Thirteen of 64 patients had a minimal iliofemoral artery luminal diameter (MLD) of ≤5.0 mm (mean 4.38+/−0.59 mm, range 3.1–5.0 mm), with vessel calcification ≤90° to 360° and tortuosity <45° to >90°. At the MLD point, the sheath-to-artery ratios, based on the fully expanded 7.67 mm OD, ranged 1.53–2.47, higher than previously reported ratios that risk vascular complications. Major comorbidities included chronic renal failure, severe chronic obstructive pulmonary disease, extreme thrombocytopenia, cirrhosis, prior cardiac surgery, poor ventricular function, and frailty. All 64 patients had TAVR with IV sedation and local anesthesia, with 0% sheath malfunction, 0% vascular complications, and 0% bleeding in-hospital and at 30 days per VARC-2 definitions. Conclusions: TAVR using a balloon-expandable, re-collapsible sheath is safe, including in small iliofemoral access ≤5.0 mm, thus considerably expanding the population suitable for transfemoral approach.
KW - TAVR
KW - VARC-2
KW - solopath
KW - transfemoral
KW - vascular complication
UR - http://www.scopus.com/inward/record.url?scp=84978786234&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84978786234&partnerID=8YFLogxK
U2 - 10.1002/ccd.26336
DO - 10.1002/ccd.26336
M3 - Article
C2 - 26603266
AN - SCOPUS:84978786234
VL - 88
SP - 135
EP - 143
JO - Catheterization and Cardiovascular Interventions
JF - Catheterization and Cardiovascular Interventions
SN - 1522-1946
IS - 1
ER -