T1 - Validating common carotid stenosis by duplex ultrasound with carotid angiogram or computed tomography scan
AU - Matos, Jesus M.
AU - Barshes, Neal R.
AU - McCoy, Sally
AU - Pisimisis, George
AU - Felkai, Deborah
AU - Kougias, Panos
AU - Lin, Peter H.
AU - Bechara, Carlos F.
N1 - Copyright:
Copyright 2014 Elsevier B.V., All rights reserved.
PY - 2014/2
Y1 - 2014/2
N2 - Background No consensus exists for duplex ultrasound criteria in the diagnosis of significant common carotid artery (CCA) stenosis. In general, peak systolic velocity (PSV) >150 cm/s with poststenotic turbulence indicates a stenosis >50%. The purpose of our study is to correlate CCA duplex velocities with angiographic findings of significant stenosis >60%. Methods We reviewed the carotid duplex records from 2008 to 2011 looking for patients with isolated CCA stenosis and no ipsilateral internal or contralateral carotid artery disease who received either a carotid angiogram or a computed tomography scan. We identified 25 patients who had significant CCA disease >60%. We also selected 74 controls without known CCA stenosis. We performed receiver operating characteristics analysis to correlate PSV and end-diastolic velocity (EDV) with angiographic stenosis >60%. The degree of stenosis was determined by measuring the luminal stenosis in comparison to the proximal normal CCA diameter. Results Most patients had a carotid angiogram (21/25), four only had a computed tomography angiography and four had both. Eighteen patients had history of neck radiation. The CCA PSV ≥250 cm/s had a sensitivity of 98.7% (81.5%-100%) and a specificity of 95.7% (92.0%-99.9%), CCA PSV ≥300 cm/s had a sensitivity of 90.9% (69.4%-98.4%) and a specificity of 98.7% (92.0%-99.9%). The CCA EDV ≥40 cm/s had a sensitivity of 95.5% (95% confidence interval of 75.1-99.8%) and specificity of 98.7% (92.0%-99.9%), EDV ≥60 cm/s had a sensitivity of 100% (75.1%-99.8%) and specificity of 87% (94.1-100%), and EDV ≥70 cm/s had a sensitivity of 86.4% (64.0%-96.4%) and specificity of 100% (94.1%-100%). The presence of both PSV <250 cm/s and EDV <60 cm/s had a 98.7% negative predictive value, and the presence of both PSV ≥250 cm/s and EDV ≥60 cm/s had 100% positive predictive value. Conclusions Establishing CCA duplex criteria to screen patients with significant stenosis is crucial to identify those who will need further imaging modality or treatment. In our laboratory, CCA PSV ≥250 cm/s and EDV ≥60 cm/s are thresholds that can be used to identify significant (>60%) CCA stenosis with a high degree of accuracy.
AB - Background No consensus exists for duplex ultrasound criteria in the diagnosis of significant common carotid artery (CCA) stenosis. In general, peak systolic velocity (PSV) >150 cm/s with poststenotic turbulence indicates a stenosis >50%. The purpose of our study is to correlate CCA duplex velocities with angiographic findings of significant stenosis >60%. Methods We reviewed the carotid duplex records from 2008 to 2011 looking for patients with isolated CCA stenosis and no ipsilateral internal or contralateral carotid artery disease who received either a carotid angiogram or a computed tomography scan. We identified 25 patients who had significant CCA disease >60%. We also selected 74 controls without known CCA stenosis. We performed receiver operating characteristics analysis to correlate PSV and end-diastolic velocity (EDV) with angiographic stenosis >60%. The degree of stenosis was determined by measuring the luminal stenosis in comparison to the proximal normal CCA diameter. Results Most patients had a carotid angiogram (21/25), four only had a computed tomography angiography and four had both. Eighteen patients had history of neck radiation. The CCA PSV ≥250 cm/s had a sensitivity of 98.7% (81.5%-100%) and a specificity of 95.7% (92.0%-99.9%), CCA PSV ≥300 cm/s had a sensitivity of 90.9% (69.4%-98.4%) and a specificity of 98.7% (92.0%-99.9%). The CCA EDV ≥40 cm/s had a sensitivity of 95.5% (95% confidence interval of 75.1-99.8%) and specificity of 98.7% (92.0%-99.9%), EDV ≥60 cm/s had a sensitivity of 100% (75.1%-99.8%) and specificity of 87% (94.1-100%), and EDV ≥70 cm/s had a sensitivity of 86.4% (64.0%-96.4%) and specificity of 100% (94.1%-100%). The presence of both PSV <250 cm/s and EDV <60 cm/s had a 98.7% negative predictive value, and the presence of both PSV ≥250 cm/s and EDV ≥60 cm/s had 100% positive predictive value. Conclusions Establishing CCA duplex criteria to screen patients with significant stenosis is crucial to identify those who will need further imaging modality or treatment. In our laboratory, CCA PSV ≥250 cm/s and EDV ≥60 cm/s are thresholds that can be used to identify significant (>60%) CCA stenosis with a high degree of accuracy.
UR - http://www.scopus.com/inward/record.url?scp=84892921518&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84892921518&partnerID=8YFLogxK
U2 - 10.1016/j.jvs.2013.08.030
DO - 10.1016/j.jvs.2013.08.030
M3 - Article
C2 - 24080127
AN - SCOPUS:84892921518
SN - 0741-5214
VL - 59
SP - 435
EP - 439
JO - Journal of Vascular Surgery
JF - Journal of Vascular Surgery
IS - 2
ER -