Thirty-day outcomes of a novel biomimetic balloon-expandable transcatheter heart valve in patients with small aortic annuli

Ole De Backer, Arif A. Khokhar, Gintautas Bieliauskas, Azeem Latib, Rishi Puri, Tamaz Shaburishvili, Maia Zhividze, Teona Zirakashvili, Anita Asgar, Amar Krishnaswamy, Santiago A. Garcia, Rebecca T. Hahn, Paul D. Mahoney, Thomas Waggoner, Gorav Ailawadi, Stanley Chetcuti, Won Keun Kim, João Cavalcante, Kari Feldt, Jeffrey J. PopmaAugustin Coisne, David Meier, Christopher U. Meduri, Stephan Windecker, Michael J. Reardon, Vinayak N. Bapat

Research output: Contribution to journalArticlepeer-review

Abstract

BACKGROUND: Transcatheter aortic valve implantation (TAVI) in patients with small aortic annuli (SAA) is associated with an increased risk of prosthesis-patient mismatch (PPM). AIMS: This study assesses the 30-day performance of the novel balloon-expandable DurAVR transcatheter heart valve (THV), which features a unique single-piece biomimetic leaflet design, in patients with SAA. METHODS: This pooled analysis derived from first-in-human and early feasibility studies includes all patients with SAA (defined as an aortic annular area from 346 mm2 to 452 mm2) treated with the small-sized DurAVR THV. The mean computed tomography (CT)-derived aortic annulus area was 404±37 mm2, with a mean diameter of 22.7±1.0 mm. Outcomes at 30 days, including PPM, were evaluated per Valve Academic Research Consortium 3 criteria, with independent adjudication of clinical events and core laboratory analysis of post-implant transthoracic echocardiograms. RESULTS: Amongst 100 patients (mean age 77.0±7.3 years; 78% female; mean Society of Thoracic Surgeons score 4.7±4.0%) treated with the DurAVR THV, the overall technical success rate was 93%. At 30 days, device success was achieved in 91% of patients, with no reported deaths and a stroke rate of 2%. Echocardiographic haemodynamic assessment showed a mean transprosthetic gradient of 8.2±3.1 mmHg, a mean effective orifice area of 2.2±0.3 cm2, and a Doppler velocity index of 0.60±0.10. The incidence of moderate or greater PPM was 3%, and no patients experienced more than mild paravalvular leak. The rate of new permanent pacemaker implantation was 6%. CONCLUSIONS: In patients with SAA, the DurAVR THV demonstrated promising clinical and echocardiographic outcomes at 30 days. Longer-term follow-up in larger cohorts is needed to confirm these encouraging early results.

Original languageEnglish (US)
Pages (from-to)e150-e160
JournalEuroIntervention
Volume22
Issue number3
DOIs
StatePublished - Feb 2 2026

Keywords

  • biomimetic leaflets
  • early outcomes
  • small annulus
  • transcatheter aortic valve

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Thirty-day outcomes of a novel biomimetic balloon-expandable transcatheter heart valve in patients with small aortic annuli'. Together they form a unique fingerprint.

Cite this