T1 - Quality assurance for Gamma Knife Perfexion using the Exradin W1 plastic scintillation detector and Lucy phantom
AU - Therriault-Proulx, Francois
AU - Pino, Ramiro
AU - Yang, James N.
AU - Sam Beddar, A.
N1 - Publisher Copyright:
© 2019 Institute of Physics and Engineering in Medicine.
PY - 2019/11/15
Y1 - 2019/11/15
N2 - The goal of this work is to validate the use of the Exradin W1 plastic scintillation detector (PSD) to measure profiles and output factors from Gamma Knife Perfexion collimators in a Lucy phantom. The Exradin W1 PSD has a small-volume, near-water-equivalent, energy-independent sensitive element. Output measurements were performed for all 3 collimators (4 mm, 8 mm, and 16 mm) of the Gamma Knife Perfexion system, and these measurements were compared to measurements made with an A16 ion chamber and an EBT3 film and to the nominal values. We showed that a configuration in which the focus or 'shot' moves while the detector remains fixed is essentially equivalent to a configuration in which the focus is fixed while the detector moves. A Lucy phantom containing a PSD was moved in small steps to acquire profiles in all three dimensions. EBT3 film was inserted in the Lucy phantom and exposed to a single shot for each collimator. The relative values for output factors measured with the PSD were 1.000, 0.892, and 0.795, for the 16 mm, 8 mm, and 4 mm collimators, respectively. The values measured with EBT3 film were 1.000, 0.881, and 0.793, and the values measured with the A16 ion chamber were 1.000, 0.883, and 0.727. The nominal output factors for the Gamma Knife Perfexion are 1.000, 0.900, and 0.814, respectively. There was excellent agreement between all profiles measured with the PSD and EBT3 as well as with the treatment planning system data provided by the vendor. In light of our results, the Exradin W1 PSD is well suited for beam quality assurance of a Gamma Knife Perfexion irradiator.
AB - The goal of this work is to validate the use of the Exradin W1 plastic scintillation detector (PSD) to measure profiles and output factors from Gamma Knife Perfexion collimators in a Lucy phantom. The Exradin W1 PSD has a small-volume, near-water-equivalent, energy-independent sensitive element. Output measurements were performed for all 3 collimators (4 mm, 8 mm, and 16 mm) of the Gamma Knife Perfexion system, and these measurements were compared to measurements made with an A16 ion chamber and an EBT3 film and to the nominal values. We showed that a configuration in which the focus or 'shot' moves while the detector remains fixed is essentially equivalent to a configuration in which the focus is fixed while the detector moves. A Lucy phantom containing a PSD was moved in small steps to acquire profiles in all three dimensions. EBT3 film was inserted in the Lucy phantom and exposed to a single shot for each collimator. The relative values for output factors measured with the PSD were 1.000, 0.892, and 0.795, for the 16 mm, 8 mm, and 4 mm collimators, respectively. The values measured with EBT3 film were 1.000, 0.881, and 0.793, and the values measured with the A16 ion chamber were 1.000, 0.883, and 0.727. The nominal output factors for the Gamma Knife Perfexion are 1.000, 0.900, and 0.814, respectively. There was excellent agreement between all profiles measured with the PSD and EBT3 as well as with the treatment planning system data provided by the vendor. In light of our results, the Exradin W1 PSD is well suited for beam quality assurance of a Gamma Knife Perfexion irradiator.
KW - Exradin W1 Scintillaror
KW - Gamma Knife
KW - output factors
KW - radiosurgery commissioning
KW - scintillation detectors
UR - http://www.scopus.com/inward/record.url?scp=85075090448&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85075090448&partnerID=8YFLogxK
U2 - 10.1088/1361-6560/ab4ac3
DO - 10.1088/1361-6560/ab4ac3
M3 - Article
C2 - 31581139
AN - SCOPUS:85075090448
SN - 0031-9155
VL - 64
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 22
M1 - 225007
ER -