TY - JOUR
T1 - Prospective study of a single-incision sling versus a transobturator sling in women with stress urinary incontinence
T2 - 3-year results
AU - White, Amanda B.
AU - Kahn, Bruce S.
AU - Gonzalez, Ricardo R.
AU - Rosamilia, Anna
AU - Anger, Jennifer T.
AU - Eilber, Karyn S.
AU - Schaffer, Joseph I.
N1 - Funding Information:
This study was sponsored by Boston Scientific; the sponsor assembled a team of collaborators to design this study in accordance with the Food and Drug Administration 522 mandate and provided statistical analysis of data.
Publisher Copyright:
© 2020 Elsevier Inc.
PY - 2020/10
Y1 - 2020/10
N2 - Background: Long-term safety and efficacy data on use of single-incision slings in stress urinary incontinence are limited. Objective: To determine whether the single-incision sling Solyx (Boston Scientific, Marlborough, MA) is noninferior to the transobturator sling Obtryx II (Boston Scientific) in efficacy and safety for treatment of stress urinary incontinence. This 522 post-market surveillance study has been designed in response to a Food and Drug Administration request to evaluate improvement in stress urinary incontinence at 36 months following single-incision sling compared with baseline, as well as provide an assessment of mesh-related complications and subject-reported outcomes, relative to the transobturator sling control. Study Design: This prospective, nonrandomized, parallel cohort, multicenter postapproval study enrolled subjects to receive single-incision sling or transobturator sling. Study sites were assigned to a cohort group based on documented competency with the cohort device. Patient follow-up was 36 months to compare efficacy and adverse events for noninferiority. Inclusion criteria included stress predominant urinary incontinence, a positive cough stress test, and post-void residual ≤150 cc. Participants were ineligible if they had undergone previous stress urinary incontinence surgery or had a previous mesh complication. Primary endpoint was treatment success defined by composite negative cough stress test and subjective improvement in stress urinary incontinence using Patient Global Impression of Improvement at 36 months. Secondary endpoints included adverse events and indications for retreatment. Noninferiority margins of 15% and 10% were prespecified for the primary efficacy and safety endpoints. Data analysis was performed using intent-to-treat and per-protocol methods. Due to the observational nature of the study, a propensity score methodology was applied to account for differences in patient and surgeon characteristics between treatment groups. The study design and variables to be included in the propensity score model were reviewed and approved by Food and Drug Administration reviewers before outcome analyses were performed. Results: No evidence of imbalance in baseline characteristics was observed between groups after propensity score stratification in the 281 subjects. EBL (72.3±92 vs 73.1±63.9 mL, P=.786), time to spontaneous void (1.1±2 vs 0.8±2.8 days, P=.998), and time to discharge (0.7±0.7 vs 0.6±0.6 days, P=.524) were similar between groups. At 36 months, treatment success was 90.4% in the single-incision sling group and 88.9% in the transobturator sling group (P=.93). At 36 months, mesh-related complications were similar between groups (mesh exposure: 2.8% vs 5.0%, P=.38). Serious adverse events including pain during intercourse (0.7% vs 0%, P=1.00), pelvic pain (0.7% vs 0%, P=1.00), and urinary retention (2.8% vs 4.3%, P=.54) were similar between groups. Conclusion: Single-incision sling was not inferior to transobturator sling for long-term treatment success of stress urinary incontinence. The rates of serious adverse events were acceptably low and similar between groups.
AB - Background: Long-term safety and efficacy data on use of single-incision slings in stress urinary incontinence are limited. Objective: To determine whether the single-incision sling Solyx (Boston Scientific, Marlborough, MA) is noninferior to the transobturator sling Obtryx II (Boston Scientific) in efficacy and safety for treatment of stress urinary incontinence. This 522 post-market surveillance study has been designed in response to a Food and Drug Administration request to evaluate improvement in stress urinary incontinence at 36 months following single-incision sling compared with baseline, as well as provide an assessment of mesh-related complications and subject-reported outcomes, relative to the transobturator sling control. Study Design: This prospective, nonrandomized, parallel cohort, multicenter postapproval study enrolled subjects to receive single-incision sling or transobturator sling. Study sites were assigned to a cohort group based on documented competency with the cohort device. Patient follow-up was 36 months to compare efficacy and adverse events for noninferiority. Inclusion criteria included stress predominant urinary incontinence, a positive cough stress test, and post-void residual ≤150 cc. Participants were ineligible if they had undergone previous stress urinary incontinence surgery or had a previous mesh complication. Primary endpoint was treatment success defined by composite negative cough stress test and subjective improvement in stress urinary incontinence using Patient Global Impression of Improvement at 36 months. Secondary endpoints included adverse events and indications for retreatment. Noninferiority margins of 15% and 10% were prespecified for the primary efficacy and safety endpoints. Data analysis was performed using intent-to-treat and per-protocol methods. Due to the observational nature of the study, a propensity score methodology was applied to account for differences in patient and surgeon characteristics between treatment groups. The study design and variables to be included in the propensity score model were reviewed and approved by Food and Drug Administration reviewers before outcome analyses were performed. Results: No evidence of imbalance in baseline characteristics was observed between groups after propensity score stratification in the 281 subjects. EBL (72.3±92 vs 73.1±63.9 mL, P=.786), time to spontaneous void (1.1±2 vs 0.8±2.8 days, P=.998), and time to discharge (0.7±0.7 vs 0.6±0.6 days, P=.524) were similar between groups. At 36 months, treatment success was 90.4% in the single-incision sling group and 88.9% in the transobturator sling group (P=.93). At 36 months, mesh-related complications were similar between groups (mesh exposure: 2.8% vs 5.0%, P=.38). Serious adverse events including pain during intercourse (0.7% vs 0%, P=1.00), pelvic pain (0.7% vs 0%, P=1.00), and urinary retention (2.8% vs 4.3%, P=.54) were similar between groups. Conclusion: Single-incision sling was not inferior to transobturator sling for long-term treatment success of stress urinary incontinence. The rates of serious adverse events were acceptably low and similar between groups.
KW - Obtryx II
KW - Solyx
KW - midurethral sling
KW - single-incision sling
KW - stress urinary incontinence
KW - transobturator sling
UR - http://www.scopus.com/inward/record.url?scp=85083842642&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85083842642&partnerID=8YFLogxK
U2 - 10.1016/j.ajog.2020.03.008
DO - 10.1016/j.ajog.2020.03.008
M3 - Article
C2 - 32184149
AN - SCOPUS:85083842642
SN - 0002-9378
VL - 223
SP - 545.e1-545.e11
JO - American Journal of Obstetrics and Gynecology
JF - American Journal of Obstetrics and Gynecology
IS - 4
ER -