TY - JOUR
T1 - Performance Assessment of a Radiofrequency Powered Guidewire for Crossing Peripheral Arterial Occlusions Based on Lesion Morphology
AU - Tavallaei, Mohammad A.
AU - Zhou, James J.
AU - Roy, Trisha L.
AU - Wright, Graham A.
N1 - Publisher Copyright:
© 2018, Biomedical Engineering Society.
PY - 2018/7/1
Y1 - 2018/7/1
N2 - Endovascular wires and devices for peripheral arterial disease therapy have evolved greatly, yet failure rates of these procedures remain high. Information on lesion composition may inform device selection to improve the success rates of these procedures. This paper, presents an approach for informed guidewire selection. The objective of this study is to quantitatively assess the performance of a radiofrequency powered guidewire in the crossing of various morphology types of peripheral chronic total occlusions. Samples taken from amputated patient limbs are characterized by magnetic resonance imaging. Using a customized catheter test station, the performance of a radiofrequency powered guidewire in puncturing these lesions is compared to a conventional guidewire, and to itself when not powered. The analysis includes quantitative and statistical comparisons of the puncture forces experienced by the different guidewires in “hard” vs. “soft” lesions as well as qualitative assessment of deflections, buckling and puncture success of the wires. Results indicate that the use of radiofrequency ablation significantly reduces the required puncture force, reduced events of buckling and deflection, and resulted in a significantly higher puncture success rate.
AB - Endovascular wires and devices for peripheral arterial disease therapy have evolved greatly, yet failure rates of these procedures remain high. Information on lesion composition may inform device selection to improve the success rates of these procedures. This paper, presents an approach for informed guidewire selection. The objective of this study is to quantitatively assess the performance of a radiofrequency powered guidewire in the crossing of various morphology types of peripheral chronic total occlusions. Samples taken from amputated patient limbs are characterized by magnetic resonance imaging. Using a customized catheter test station, the performance of a radiofrequency powered guidewire in puncturing these lesions is compared to a conventional guidewire, and to itself when not powered. The analysis includes quantitative and statistical comparisons of the puncture forces experienced by the different guidewires in “hard” vs. “soft” lesions as well as qualitative assessment of deflections, buckling and puncture success of the wires. Results indicate that the use of radiofrequency ablation significantly reduces the required puncture force, reduced events of buckling and deflection, and resulted in a significantly higher puncture success rate.
KW - Chronic total occlusions
KW - Endovascular interventions
KW - Guidewires
KW - Magnetic resonance imaging
KW - Peripheral arterial disease
KW - Radio frequency ablation
UR - http://www.scopus.com/inward/record.url?scp=85044785408&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85044785408&partnerID=8YFLogxK
U2 - 10.1007/s10439-018-2021-y
DO - 10.1007/s10439-018-2021-y
M3 - Article
C2 - 29616386
AN - SCOPUS:85044785408
SN - 0090-6964
VL - 46
SP - 940
EP - 946
JO - Annals of Biomedical Engineering
JF - Annals of Biomedical Engineering
IS - 7
ER -