Long-term in vivo alterations of polyester vascular grafts in humans

G. Riepe, J. Loos, H. Imig, A. Schroder, E. Schneider, J. Petermann, A. Rogge, M. Ludwig, A. Schenke, R. Nassutt, N. Chakfe, M. Morlock

    Research output: Contribution to journalArticlepeer-review

    112 Scopus citations

    Abstract

    Objectives: To examine the influence of in vivo hydrolysis on the physical properties of polyester grafts and their correlation to the period of implantation in the human body. Materials and methods: Sixty-five explanted vascular grafts were obtained after 0-23 years of implantation due to suture aneurysms (18), occlusion (12), graft infection (12), failure of graft material (7) and post-mortem (16). The surface was examined by scanning electron microscopy, the molecular integrity by infra-red spectroscopy and physical strength by probe puncture. Results: Scission of macromolecular chains and loss of strength were shown. It was demonstrated that hydrolytic degradation of polyester takes place with increasing time of implantation in humans. Analysis by linear regression showed that polyester grafts lose 31.4% of their bursting strength in 10 years and 100% in 25-39 years after implantation. Conclusions: Regular follow-ups of patients with aged vascular grafts and the precise documentation of implanted materials are necessary to estimate graft degradation.

    Original languageEnglish (US)
    Pages (from-to)540-548
    Number of pages9
    JournalEuropean Journal of Vascular and Endovascular Surgery
    Volume13
    Issue number6
    DOIs
    StatePublished - 1997

    Keywords

    • Degradation
    • Hydrolysis
    • Polyester
    • Vascular grafts

    ASJC Scopus subject areas

    • Surgery
    • Cardiology and Cardiovascular Medicine

    Fingerprint

    Dive into the research topics of 'Long-term in vivo alterations of polyester vascular grafts in humans'. Together they form a unique fingerprint.