Inappropriate implantable cardioverter-defibrillator shocks: Attributed to alternating-current leak in a swimming pool

John N. Makaryus, Julia Angert-Gilman, Mena Yacoub, Apoor Patel, Bruce Goldner

Research output: Contribution to journalArticlepeer-review

5 Scopus citations

Abstract

Implantable cardioverter-defibrillators (ICDs) are the standard of care for preventing sudden cardiac death in patients who are predisposed to malignant ventricular arrhythmias. Causes of inappropriate ICD shock include equipment malfunction, improper arrhythmia evaluation, misinterpretation of myopotentials, and electromagnetic interference. As the number of implanted ICDs has increased, other contributors to inappropriate therapy have become known, such as minimal electrical current leaks that mimic ventricular fibrillation. We present the case of a 63-year-old man with a biventricular ICD who received 2 inappropriate shocks, probably attributable to alternating-current leaks in a swimming pool. In addition, we discuss ICD sensitivity and offer recommendations to avoid similar occurrences.

Original languageEnglish (US)
Pages (from-to)61-63
Number of pages3
JournalTexas Heart Institute Journal
Volume41
Issue number1
DOIs
StatePublished - Feb 2014

Keywords

  • Arrhythmias
  • Artificial
  • Cardiac pacing
  • Cardiac/etiology
  • Defibrillators
  • Electricity/adverse effects
  • Electromagnetic phenomena
  • Electrophysiology
  • Environmental exposure
  • Equipment safety
  • Implantable/adverse effects
  • Signal-to-noise ratio
  • Swimming pools

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Inappropriate implantable cardioverter-defibrillator shocks: Attributed to alternating-current leak in a swimming pool'. Together they form a unique fingerprint.

Cite this