TY - JOUR
T1 - Imaging of the Left Atrial Appendage before Occluder Device Placement
T2 - Evaluation of Virtual Monoenergetic Images in a Single-Bolus Dual-Phase Protocol
AU - Laukamp, Kai Roman
AU - Dastmalchian, Sara
AU - Tandon, Yasmeen K.
AU - Ciancibello, Leslie
AU - Pennig, Lenhard
AU - Lennartz, Simon
AU - Al-Kindi, Sadeer
AU - Rajagopalan, Sanjay
AU - Bera, Kaustav
AU - Hokamp, Nils Große
AU - Gilkeson, Robert
AU - Gupta, Amit
N1 - Publisher Copyright:
© Wolters Kluwer Health, Inc. All rights reserved.
PY - 2022/9/1
Y1 - 2022/9/1
N2 - Purpose Preimplantation cardiac computed tomography (CT) for assessment of the left atrial appendage (LAA) enables correct sizing of the device and the detection of contraindications, such as thrombi. In the arterial phase, distinction between false filling defects and true thrombi can be hampered by insufficient contrast medium distribution. A delayed scan can be used to further differentiate both conditions, but contrast in these acquisitions is relatively lower. In this study, we investigated whether virtual monoenergetic images (VMI) from dual-energy spectral detector CT (SDCT) can be used to enhance contrast and visualization in the delayed phase. Materials and Methods Forty-nine patients receiving SDCT imaging of the LAA were retrospectively enrolled. The imaging protocol comprised dual-phase acquisitions with single-bolus contrast injection. Conventional images (CI) from both phases and 40-keV VMI from the delayed phase were reconstructed. Attenuation, signal-, and contrast-to-noise ratios (SNR/CNR) were calculated by placing regions-of-interest in the LAA, left atrium, and muscular portion of interventricular septum. Two radiologists subjectively evaluated conspicuity and homogeneity of contrast distribution within the LAA. Results Contrast of the LAA decreased significantly in the delayed phase but was significantly improved by VMI, showing comparable attenuation, SNR, and CNR to CI from the arterial phase (attenuation/SNR/CNR, CI arterial phase: 266.0 ± 117.0 HU/14.2 ± 7.2/6.6 ± 3.9; CI-delayed phase: 107.6 ± 35.0 HU/5.9 ± 3.0/1.0 ± 1.0; VMI delayed phase: 260.3 ± 108.6 HU/18.2 ± 10.6/4.8 ± 3.4). The subjective reading confirmed the objective findings showing improved conspicuity and homogeneity in the delayed phase. Conclusions The investigated single-bolus dual-phase acquisition protocol provided improved visualization of the LAA. Homogeneity of contrast media was higher in the delayed phase, while VMI maintained high contrast.
AB - Purpose Preimplantation cardiac computed tomography (CT) for assessment of the left atrial appendage (LAA) enables correct sizing of the device and the detection of contraindications, such as thrombi. In the arterial phase, distinction between false filling defects and true thrombi can be hampered by insufficient contrast medium distribution. A delayed scan can be used to further differentiate both conditions, but contrast in these acquisitions is relatively lower. In this study, we investigated whether virtual monoenergetic images (VMI) from dual-energy spectral detector CT (SDCT) can be used to enhance contrast and visualization in the delayed phase. Materials and Methods Forty-nine patients receiving SDCT imaging of the LAA were retrospectively enrolled. The imaging protocol comprised dual-phase acquisitions with single-bolus contrast injection. Conventional images (CI) from both phases and 40-keV VMI from the delayed phase were reconstructed. Attenuation, signal-, and contrast-to-noise ratios (SNR/CNR) were calculated by placing regions-of-interest in the LAA, left atrium, and muscular portion of interventricular septum. Two radiologists subjectively evaluated conspicuity and homogeneity of contrast distribution within the LAA. Results Contrast of the LAA decreased significantly in the delayed phase but was significantly improved by VMI, showing comparable attenuation, SNR, and CNR to CI from the arterial phase (attenuation/SNR/CNR, CI arterial phase: 266.0 ± 117.0 HU/14.2 ± 7.2/6.6 ± 3.9; CI-delayed phase: 107.6 ± 35.0 HU/5.9 ± 3.0/1.0 ± 1.0; VMI delayed phase: 260.3 ± 108.6 HU/18.2 ± 10.6/4.8 ± 3.4). The subjective reading confirmed the objective findings showing improved conspicuity and homogeneity in the delayed phase. Conclusions The investigated single-bolus dual-phase acquisition protocol provided improved visualization of the LAA. Homogeneity of contrast media was higher in the delayed phase, while VMI maintained high contrast.
KW - dual-energy CT
KW - left atrial appendage
KW - occluder device
KW - spectral detector
KW - x-ray computed tomography
UR - http://www.scopus.com/inward/record.url?scp=85138449206&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85138449206&partnerID=8YFLogxK
U2 - 10.1097/RCT.0000000000001330
DO - 10.1097/RCT.0000000000001330
M3 - Article
C2 - 35723620
AN - SCOPUS:85138449206
SN - 0363-8715
VL - 46
SP - 735
EP - 741
JO - Journal of Computer Assisted Tomography
JF - Journal of Computer Assisted Tomography
IS - 5
ER -