N2 - Objective. Direct-current electrocorticography (DC-ECoG) allows a more complete characterization of brain states and pathologies than traditional alternating-current recordings (AC-ECoG). However, reliable recording of DC signals is challenging because of electrode polarization-induced potential drift, particularly at low frequencies and for more conducting materials. Further challenges arise as electrode size decreases, since impedance is increased and the potential drift is augmented. While microelectrodes have been investigated for AC-ECoG recordings, little work has addressed microelectrode properties for DC-signal recording. In this paper, we investigated several common microelectrode materials used in biomedical application for DC-ECoG. Approach. Five of the most common materials including gold (Au), silver/silver chloride (Ag/AgCl), platinum (Pt), Iridium oxide (IrOx), and platinum-iridium oxide (Pt/IrOx) were investigated for electrode diameters of 300 μm. The critical characteristics such as polarization impedance, AC current-induced polarization, long-term stability and low-frequency noise were studied in vitro (0.9% saline). The two most promising materials, Pt and Pt/lrOx were further investigated in vivo by recording waves of spreading depolarization, one of the most important applications for DC-ECoG in clinical and basic science research. Main results. Our experimental results indicate that IrOx-based microelectrodes, particularly with composite layers of nanostructures, are excellent in all of the common evaluation characteristics both in vitro and in vivo and are most suitable for multimodal monitoring applications. Pt electrodes suffer high current-induced polarization, but have acceptable long-term stability suitable for DC-ECoG. Major significance. The results of this study provide quantitative data on the electrical properties of microelectrodes with commonly-used materials and will be valuable for development of neural recordings inclusive of low frequencies.
AB - Objective. Direct-current electrocorticography (DC-ECoG) allows a more complete characterization of brain states and pathologies than traditional alternating-current recordings (AC-ECoG). However, reliable recording of DC signals is challenging because of electrode polarization-induced potential drift, particularly at low frequencies and for more conducting materials. Further challenges arise as electrode size decreases, since impedance is increased and the potential drift is augmented. While microelectrodes have been investigated for AC-ECoG recordings, little work has addressed microelectrode properties for DC-signal recording. In this paper, we investigated several common microelectrode materials used in biomedical application for DC-ECoG. Approach. Five of the most common materials including gold (Au), silver/silver chloride (Ag/AgCl), platinum (Pt), Iridium oxide (IrOx), and platinum-iridium oxide (Pt/IrOx) were investigated for electrode diameters of 300 μm. The critical characteristics such as polarization impedance, AC current-induced polarization, long-term stability and low-frequency noise were studied in vitro (0.9% saline). The two most promising materials, Pt and Pt/lrOx were further investigated in vivo by recording waves of spreading depolarization, one of the most important applications for DC-ECoG in clinical and basic science research. Main results. Our experimental results indicate that IrOx-based microelectrodes, particularly with composite layers of nanostructures, are excellent in all of the common evaluation characteristics both in vitro and in vivo and are most suitable for multimodal monitoring applications. Pt electrodes suffer high current-induced polarization, but have acceptable long-term stability suitable for DC-ECoG. Major significance. The results of this study provide quantitative data on the electrical properties of microelectrodes with commonly-used materials and will be valuable for development of neural recordings inclusive of low frequencies.
KW - DC-ECoG
KW - ECoG
KW - electrode polarization
KW - low-frequency noise
KW - microelectrode
KW - polarization impedance
KW - spreading depolarization
UR - http://www.scopus.com/inward/record.url?scp=84957812498&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84957812498&partnerID=8YFLogxK
U2 - 10.1088/1741-2560/13/1/016008
DO - 10.1088/1741-2560/13/1/016008
M3 - Article
C2 - 26655565
AN - SCOPUS:84957812498
VL - 13
JO - Journal of neural engineering
JF - Journal of neural engineering
SN - 1741-2560
IS - 1
M1 - 016008
ER -