Estimation of thermal dose from MR thermometry during application of nonablative pulsed high intensity focused ultrasound

Brian E. O'Neill, Christof Karmonik, Elisabetta Sassaroli, King C. Li

Research output: Contribution to journalArticlepeer-review

16 Scopus citations

Abstract

Purpose: To evaluate whether MR thermometry is sufficiently fast, accurate, and spatially resolved for monitoring the thermal safety of nonablative pulsed high intensity ultrasound (pHIFU) treatments. Materials and Methods: A combination of real MR thermometry data and modeling was used to analyze the effects of temporal and spatial averaging as well as noise on the peak temperatures and thermal doses that would be measured by MR thermometry. Results: MR thermometry systematically underestimates the temperature and thermal doses during pHIFU treatment. Small underestimates of peak temperature can lead to large underestimates of thermal dose. Spatial averaging errors are small for ratios of pixel dimension to heating zone radius less than 0.25, which may be achieved by reducing the voxel size or steering the acoustic beam. Thermal dose might also be underestimated for very short, high power pulses due to temporal averaging. A simple correction factor based on the applied power and duty cycle may be applied to determine the upper bound of this effect. Conclusion: The temperature and thermal dose measured using MR thermometry during pulsed HIFU treatment is probably sufficient in most instances. Simple corrections may be used to calculate an upper bound where this is a critical factor.

Original languageEnglish (US)
Pages (from-to)1169-1178
Number of pages10
JournalJournal of Magnetic Resonance Imaging
Volume35
Issue number5
DOIs
StatePublished - May 2012

Keywords

  • Drug delivery
  • HIFU
  • MR thermometry
  • Pulsed high intensity focused ultrasound
  • Ultrasound hyperthermia

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Estimation of thermal dose from MR thermometry during application of nonablative pulsed high intensity focused ultrasound'. Together they form a unique fingerprint.

Cite this