Abstract
OBJECTIVE: To determine whether repeated use weakens titanium aneurysm clips, making them unsafe for clinical use. METHODS: Nine Yaşargil (titanium alloy) and five Spetzler (pure titanium) clips were subjected to clinically relevant maneuvers, and the effects on the closing force were assessed. Clips were considered unsafe if 1) the blades crossed, 2) the clips failed to close completely, or 3) the blood pressure could open them. RESULTS: Yaşargil and Spetzler clips significantly (P < 0.01 for both) weakened, by 18.9% and 26.1%, respectively, after 10 minutes of sustained maximal opening. After 100 open-close cycles, Spetzler clips showed no change in closing force, but the closing force of Yaşargil clips decreased significantly (P < 0.01), by a further 12.0%. CONCLUSION: Repeated use weakens Spetzler and Yaşargil clips. All Spetzler and all but one Yaşargil clips were deemed safe for clinical use at the end of the experiments.
Original language | English (US) |
---|---|
Pages (from-to) | 966-972 |
Number of pages | 7 |
Journal | Neurosurgery |
Volume | 54 |
Issue number | 4 |
DOIs | |
State | Published - Apr 2004 |
Keywords
- Aneurysm clip
- Closing force
- Opening force
- Spring fatigue
- Titanium
- Vascular neurosurgery