Effect of transcatheter aortic valve size and position on valve-in-valve hemodynamics: An in vitro study

Ali N. Azadani, Michael Reardon, Matheus Simonato, Gabriel Aldea, Georg Nickenig, Ran Kornowski, Danny Dvir

Research output: Contribution to journalArticlepeer-review

49 Scopus citations

Abstract

Objective Transcatheter heart valve implantation in failed aortic bioprostheses (valve-in-valve [ViV]) is an increasingly used therapeutic option for high-risk patients. However, high postprocedural gradients are a significant limitation of aortic ViV. Our objective was to evaluate Medtronic CoreValve Evolut R ViV hemodynamics in relation to the degree of device oversizing and depth of implantation. Methods Evolut R devices of 23 and 26 mm were implanted within 21-, 23-, and 25-mm Hancock II bioprostheses. Small and gradual changes in implantation depth were attempted. Hemodynamic testing was performed in a pulse duplicator under ISO-5840 standard. Results A total of 47 bench-testing experiments were performed. The mean gradient of the 26-mm Evolut R in 23- and 25-mm Hancock II was lower than 23-mm Evolut R (P < .001). However, the mean gradient of 26-mm Evolut R in 21-mm Hancock II bioprostheses R (ranging from 21.30 ± 0.23 to 24.30 ± 0.22 mm Hg) was worse than 23-mm Evolut R (ranging from 15.94 ± 0.18 to 20.35 ± 0.16 mm Hg, P < .001). Furthermore, our results suggest that supra-annular implantation of 23-mm and 26-mm Evolut R devices within the bioprostheses can lead to lower gradient and improved leaflet coaptation. Regardless of implantation depth, superior transvalvular gradient is expected with 26-mm Evolut R than 23-mm Evolut R in a nonstenotic Hancock II with a true internal diameter > 17.5 mm. Conclusions The current comprehensive bench-testing assessment demonstrates the importance of both transcatheter heart valve size and device position for the attainment of optimal hemodynamics during ViV procedures. Additional in vitro testing may be required to develop hemodynamics-based guidelines for device sizing in ViV procedures in degenerated surgical bioprostheses.

Original languageEnglish (US)
Pages (from-to)1303-1315.e1
JournalJournal of Thoracic and Cardiovascular Surgery
Volume153
Issue number6
DOIs
StatePublished - Jun 2017

Keywords

  • CoreValve Evolut R
  • in vitro testing
  • transcatheter heart valve replacement
  • valve-in-valve
  • valvular hemodynamics

ASJC Scopus subject areas

  • Surgery
  • Pulmonary and Respiratory Medicine
  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Effect of transcatheter aortic valve size and position on valve-in-valve hemodynamics: An in vitro study'. Together they form a unique fingerprint.

Cite this