Abstract
The question is addressed whether patients with thoracotomy defibrillators and failing epicardial electrodes can be effectively treated with the implantation of prepectoral unipolar ('active can') defibrillators. Results indicate that abandoned epicardial patches in the pathway of unipolar defibrillation currents do not affect defibrillation thresholds and active can efficacy.
Original language | English (US) |
---|---|
Pages (from-to) | 1408-1409 |
Number of pages | 2 |
Journal | American Journal of Cardiology |
Volume | 79 |
Issue number | 10 |
DOIs | |
State | Published - May 15 1997 |
ASJC Scopus subject areas
- Cardiology and Cardiovascular Medicine