Development and validation of an impedance biosensor for point-of-care detection of vascular cell adhesion molecule-1 toward lupus diagnostics

Anjan Panneer Selvam, Andi Wangzhou, Michael Jacobs, Tianfu Wu, Chandra Mohan, Shalini Prasad

Research output: Contribution to journalArticlepeer-review

15 Scopus citations

Abstract

Aim: Systemic lupus erythematosus is an autoimmune disease that requires chronic monitoring. In this study, we demonstrate a proof-of-concept study of a highly attractive noninvasive strategy for monitoring systemic lupus erythematosus through biomarker quantification. Results: This sensor technology requires 50 μl of urine to detect and quantify vascular cell adhesion molecule-1 in 15 min. The sensor used nonfaradaic detection to demonstrate performance with and without detection antibody. Binding of immunoassay and target biomarkers were quantified with an impedance electrical immunoassay and correlated with an equivalent circuit. Conclusion: The novel sensor technology demonstrates detection in the range of 8 fg/ml to 800 pg/ml and comparative analysis with ELISA platforms was performed for 12 patient urine samples.

Original languageEnglish (US)
Article numberFSO224
JournalFuture Science OA
Volume3
Issue number3
DOIs
StatePublished - 2017

Keywords

  • VCAM-1
  • impedance biosensor
  • rapid urine screening test

ASJC Scopus subject areas

  • Biotechnology

Fingerprint

Dive into the research topics of 'Development and validation of an impedance biosensor for point-of-care detection of vascular cell adhesion molecule-1 toward lupus diagnostics'. Together they form a unique fingerprint.

Cite this