Accuracy and precision of analyses for total cholesterol as measured with the Reflotron cholesterol method

P. S. Bachorik, R. H. Bradford, T. Cole, I. Frantz, A. M. Gotto, K. Roberts, G. R. Warnick, O. D. Williams

Research output: Contribution to journalArticlepeer-review

26 Scopus citations

Abstract

We compared plasma cholesterol measurements made with Boehringer Mannheim Reflotron reflectance photometric analyzer in 1298 capillary blood samples with measurements made in venous blood samples collected at the same time and analyzed in four standardized Lipid Reserach Clinics laboratories. The Reflotron measurements averaged 0.8% to 7.8% lower than the laboratory values. Correlations (r) between the two sets of measurements ranged from 0.92 to 0.96. In some samples, however, the Reflotron values differed from the laboratory values by ≥ 12%; the cholesterol concentrations in these samples tended to be higher than in those for which better agreement was observed. The smaller negative biases were observed when test strips were used that were calibrated with reference to the Centers for Disease Control Reference Method for cholesterol. The agreement between sequential Reflotron values averaged ≤ 4.3%. There was an average difference of ≤ 1.0% between Reflotron measurements made in each of two sequential capillary blood samples taken from a single finger puncture.

Original languageEnglish (US)
Pages (from-to)1734-1739
Number of pages6
JournalClinical Chemistry
Volume35
Issue number8
DOIs
StatePublished - 1989

ASJC Scopus subject areas

  • Clinical Biochemistry
  • Biochemistry, medical

Fingerprint

Dive into the research topics of 'Accuracy and precision of analyses for total cholesterol as measured with the Reflotron cholesterol method'. Together they form a unique fingerprint.

Cite this